9 Maintenance and repair
900-301_IFU-V2.2_11316-S0-20131023-EN Operating Manual ARC 250 / 303 57
9. Maintenance and repair
Pos: 67 /679-BOW A/HF-Geräte/ ARC 300/350/9 Wartung/Wartung /10.2 Reparat ur @ 6\mod_ 12974179416 96_6.doc @ 4 8415 @ 232
9.1. Maintenance
DANGER
Infection hazard!
Carry out a surface disinfection and wrap the device in addition to
the shipping packaging material before allowing the device to leave
the hospital or office to avoid spreading germs and infections.
Check the device, the device trolley and the accessories (e.g. foot
switch, cable) after each use for damage or defects. In particular,
make sure that the insulation is intact on all cables.
Do not use any damaged device, damaged device trolley or
damaged accessories.
Replace defective accessories immediately.
Have the safety inspection for the device performed once a year.
Please consult and comply with the respective service instructions
for additional technical information.
9.1.1. Safety inspection
Safety inspections must be performed once a year.
The device and accessories may be inspected only by persons
who have the required training, knowledge or experience and who
can perform the inspection independently.
With regard to the safety inspection, you must comply with the
country-specific rules and regulations.
The inspector documents the inspection results and measured values
corresponding to the printed inspection protocol.
In the case of severe deviations from the values of the attached final
acceptance report, or if the specified maximum values were exceeded:
Send the HF device to the service center, see section Technical
service, page 59.