13 EMC
86 Operating Manual ARC 250 / 303 900-301_IFU-V2.2_11316-S0-20131023-EN
13. EMC
Pos: 77 /679-BOW A/HF-Geräte/ ARC 200/250/3 00e/13 EMV/EMV @ 6\mod_ 1297077875715_ 6.doc @ 4819 9 @ 2
13.1. Guidelines and manufacturer's declaration in
accordance with IEC 60601-1-2,
Section 6.8.3.201
Electromagnetic interference (IEC 60601-1-2, Table 201)
The ARC 250 and ARC 303 are intended to be used in an electromagnetic environment as described below.
The customer or user of the ARC 250 and ARC 303 should ensure that these devices are operated in such an
environment.
Interference measurements
Interference measurements
Interference measurements
HF emissions as per CISPR 11
The ARC 250 and ARC 303 must
emit electromagnetic energy in order
to perform their intended functions.
Nearby electronic devices may be
affected.
HF emissions as per CISPR 11
The ARC 250 and ARC 303 devices
are suitable for use in facilities other
than those suitable for a residental
environment or the like connected
directly to the public power grid,
which also supplies power to
buildings used for residential
purposes.
Emission of harmonics as per
IEC 61000-3-2
Emission of voltage fluctuations/flicker as
per IEC 61000-3-3