6 Operation
48 Operating Manual ARC 250 / 303 900-301_IFU-V2.2_11316-S0-20131023-EN
6.4.6. Menu program 4 "Show Prev Inf-No"
The errors saved in the HF device are called up in the "Show Prev Inf-
No" menu program.
1. Press the power limitation key for monopolar cutting 4.
2. Press the key for scabbing level 5/6 to call up individual errors.
The error number will appear on the main display 15.
3. Press the key/display for AUTOSTART 26 to call up the last 10
errors.
The error number will appear on the main display 15.
4. Press the power limitation key for monopolar cutting 3 to return to
the menu programs.
Pos: 55 /679-BOW A/HF-Geräte/ ARC 200/250/3 00e/6 Bedienung/M enü-Progra mm 5 "Hide Fi x Prog" @ 5\ mod_1270109 958148_6.doc @ 3 8644 @ 3
6.4.7. Menu program 5 "Hide Fix Prog"
The "Hide Fix Prog" menu program hides standard programs that are
not required.
1. Press the power limitation key for monopolar cutting 4.
2. Press the power limitation key for bipolar coagulation 24 to display
all standard programs.
The message "All FixPgm active" appears on the display/main
display 15.
3. Press the power limitation key for bipolar coagulation 23 to hide all
standard programs.
The message "No FixPgm active" appears on the display/main
display 15.
– or –
Press the key/display for AUTOSTART 26 to show or hide
individual standard programs.
4. Press program key 16/17 to browse the standard programs.
5. Press the power limitation key for monopolar coagulation 10 to hide
the standard program.
– or –
Press the power limitation key for monopolar coagulation 11 to
show the standard program.
6. Press the power limitation key for monopolar cutting 3 to return to
the menu programs.