6 Operation
40 Operating Manual ARC 250 / 303 900-301_IFU-V2.2_11316-S0-20131023-EN
6.2. Basic program settings
6.2.1. Calling up a program
Call up each of the selectable programs with the following procedure:
1. Press the program key 16/17 repeatedly until the corresponding
program is shown on the display.
2. Connect the instrruments, see section Connecting instruments,
page 35.
6.2.2. Changing the program
The preconfigured programs can be selected from 0 to 7. The
parameters of the preconfigured programs can be changed, and the new
parameters can be saved. Only the specific types of current cannot be
changed. To change the parameters, make settings such as power level,
blend factor, coagulation mode and foot switch assignment and then
store them by pressing the AUTOSTART key 26 and keeping it pressed.
Programs 8 to 99 are fully user configurable. These programs output
standard types of current.
You can change the following factory settings:
Select cutting settings
Press the power limitation key for monopolar cutting 3 or 4.
Set blend factor
1. Press key 5 or 6 for the monopolar scabbing level.
2. Set the blend factor from 1 to 9, or to 0 for coagulation-free cutting.
Select current type for coagulation current
1. Press key/display 12 for "Moderate Coagulation" current
– or –
Press key/display 13 for "Forced Coagulation" current
– or –
Press key/display 14 for "Spray Coagulation" current
2. Set the power limit for the monopolar coagulation current using the
keys for power limitation for monopolar coagulation 10/11.
– or –
1. Set the power limit for the bipolar coagulation current using the
keys for power limitation for bipolar coagulation 23/24.
Pos: 40 /679-BOW A/HF-Geräte/ ARC 200/250/3 00e/6 Bedienung/ Programm- Beschreibung en @ 5\mod_127 0108336476_ 6.doc @ 38282 @ 2