3 Functionality
900-301_IFU-V2.2_11316-S0-20131023-EN Operating Manual ARC 250 / 303 27
3.1.5. "GastroCut" mode
If you wish to change the settings for "GastroCut" mode, see
section Menu program 3A "GastroCut-Modes", page 47.
This mode consists of a pulse sequence of cutting current and
coagulation phase.
This mode is used for the removal of polyps or for making incisions in
papillae. The following applies:
Blend 0 = low coagulation rate
Blend 9 = high coagulation rate
Three sub-modes are available in "GastroCut" mode:
Mode
Characteristics
Application
Slow
Cutting characteristics:
Slower pulse sequence
Recommended for particularly intricate
tasks
Medium
Cutting characteristics:
Dynamic
Recommended for skilled users
(default setting)
Fast
Cutting characteristics:
Dynamic and fast pulse sequence
Recommended for experienced users in
this specialist area