6 Operation
900-301_IFU-V2.2_11316-S0-20131023-EN Operating Manual ARC 250 / 303 41
6.3. Program descriptions
The following recommended settings for standard programs are based
on empirical values and must be verified in each individual case by the
surgeon.
Pos: 41 /679-BOW A/HF-Geräte/ ARC 200/250/3 00e/6 Bedienung/ Programm 0 "Standard" @ 5\ mod_127010 8364476_6.doc @ 38301 @ 3
6.3.1. Program 0 "Standard"
This program is used for monopolar/bipolar cutting and coagulation in
standard applications.
Pos: 42 /679-BOW A/HF-Geräte/ ARC 200/250/3 00e/6 Bedienung/ Programm 1 "Macro" @ 6\ mod_1297072 010043_6.doc @ 4 8028 @ 3
6.3.2. Program 1 "Macro"
This program is used in plastic surgery of the hand and maxillofacial
region.
Pos: 43 /679-BOW A/HF-Geräte/ ARC 200/250/3 00e/6 Bedienung/ Programm 2 "Micro" @ 6\ mod_1297071058 574_6.doc @ 47990 @ 3
6.3.3. Program 2 "Micro"
This program is used in the lower output range with fine electrodes for
microscopic tissue structures.
Pos: 44 /679-BOW A/HF-Geräte/ ARC 200/250/3 00e/6 Bedienung/ Programm 3 "Resektion" @ 6 \mod_1297 071204215_6.doc @ 48009 @ 3
6.3.4. Program 3 "Resection"
Both cutting and coagulation are possible with bipolar
operation.
Use the socket connector for monopolar instruments 29/30
for monopolar Resection/TUR.
Only the Standard coagulation current is available at the
"Bipolar Output" 31.
This program is used for underwater cutting during transurethral and
transvaginal resection and vaporisation on prostata, bladder, and
uterus.
In this program, the resection snare is used for cutting and coagulation.
Arc control technology generates the cutting effect with simultaneously
minimized output power.
ARC control facilitates immediate cutting and prevents adhesion of the
electrodes.
The key/display for "Monopolar Output" 29/30 and "Bipolar
Output" 31 are preconfigured.
Connect the monopolar resectoscope to the "Monopolar
Output" 29/30.
Pos: 45 /679-BOW A/HF-Geräte/ ARC 200/250/3 00e/6 Bedienung/ Programm 4 "Argon" (ARC 2 00/250/300e nur mit Option G astroCut) @ 6\ mod_129707 2137808_6.doc @ 48047 @ 3