11 Technical specifications
900-301_IFU-V2.2_11316-S0-20131023-EN Operating Manual ARC 250 / 303 71
Values measured on ohmic resistances
Output power P [W] versus load
resistance R [Ω] with "Gastro KNIFE",
"Blend 0–9" setting, COAG phase
Values measured on ohmic resistances
Output power P [W] versus load
resistance R [Ω] with "Gastro KNIFE"
setting during Cut phase
Output power P [W] versus Blend
setting for "Gastro KNIFE". Rated load
resistance: Coag 75 Ω, Cut 500 Ω
HF output voltage U [Vp] versus
"Gastro KNIFE", "Blend 0–9" (no load)
The short-term peak power exceeds 400 W, but this is permissible
because the mean value over 1 second is below 400 W.