C-6
Model 0254
Section C Serial Communications Protocol Installation and Operation Manual
X-SE-0254-eng
Part Number: 541B129AAG
September, 2010
C-3-2 Command Addressing
Each Model 0254 unit is assigned a unique 5-digit address, from 0 to
65535, which must be pre-programmed in the instrument prior to
deployment in a networked system. The ports that comprise a channel are
designated with an appended [.x} or [.xx] sub-address.
The 5-digit address may be omitted when operating a single un-networked
unit, but must be used in multiple unit networks to differentiate the units
from each other.
Command arguments are single ASCII alpha non-case sensitive
characters. No spaces are allowed between the 'AZ' and <cr>, i.e. start and
respectively the end of the command.
HOST SEND
AZ yyyyy .xx K <cr>
|| || |
| | | | + — message delimiter
| | | +— command argument
| | +— port sub-address
| +— unit address
+— message prelimiter
AZ[.xx]<argument><cr> Non-networked sub-addressed port
AZ[yyyyy.xx]<argument><cr> Networked sub-addressed port
RESPONSE - <argument> dependent