EasyManua.ls Logo

Brooks 0254 - C-4-2-2 Send Channel Input Port Programmed Values

Brooks 0254
124 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C-12
Model 0254
Section C Serial Communications Protocol Installation and Operation Manual
X-SE-0254-eng
Part Number: 541B129AAG
September, 2010
C-4-2-2 Send Channel Input Port Programmed Values
This command enables a terminal operator or host to acquire the present
state of input port programmed values. Valid input port numbers (xx) are
[1, 3, 5, 7].
HOST SEND
AZ[.xx]V<cr> Non-network
AZ[yyyyy.xx]V<cr> Network
RESPONSE
PROGRAM VALUES - Channel 1 - Port 01
<04> Measure Units ml
<10> Time Base min
<03> Decimal Point x.xx
<27> Gas Factor 1.000
<28> Log Type Off
<00> PV Signal Type 0-20mA
<09> PV Full Scale 20.00 ml/m
Each of the input port programmed operating values can be individually
queried or changed in accordance with the procedures in Section C-5-1
and C-5-2.

Table of Contents