2.3 Recommended Storage Practice
If equipment supplied by Brooks Instrument is to be placed in intermediate or long-term storage,
it is recommended that it be:
a. Stored within the original shipping container.
b. Stored in a sheltered area, under the following conditions:
1) Optimum ambient temperature 21°C (70 °F), maximum 32 °C (90 °F), minimum 7 °C
(45 °F),.
2) Optimum relative humidity 45% (maximum 60%/minimum 25%).
c. Subjected to a visual inspection upon removal from storage, to verify that the condition of the
equipment is ‘as received’. If the equipment has been in storage for more than ten months or if it
has not been stored under the recommended conditions, all pressure containing seals should be
replaced. In addition, the device should be subjected to a pneumatic pressure test in accordance
with the applicable vessel codes.
2.4 Return Shipment
Prior to returning any instrument to the factory, contact your nearest Brooks location for a Return
Materials Authorization Number (RMA#). This can be obtained from one of the following
locations:
Brooks Instrument
407 W. Vine Street
P.O. Box 903
Hatfield, PA 19440 USA
Toll Free (888) 554-FLOW (3569)
Tel (215) 362-3700
Fax (215) 362-3745
E-mail: BrooksAm@BrooksInstrument.com
http://www.brooksinstrument.com
Brooks Instrument Brooks Instrument
Neonstraat 3 1-4-4 Kitasuna Koto-Ku
6718 WX Ede, Netherlands Tokyo, 136-0073 Japan
P.O. Box 428 Tel 011-81-3-5633-7100
6710 BK Ede, Netherlands Fax 011-81-3-5633-7101
Tel 31-318-549-300 E-mail: BrooksAs@BrooksInstrument.com
Fax 31-318-549-309
E-mail:BrooksEu@BrooksInstrument.com
Any instrument returned to Brooks requires completion of Form RPR003-1, Brooks Instrument
Decontamination Statement, as well as, a Material Safety Data Sheet (MSDS) for the fluid(s)
used in the instrument. This is required before any Brooks Personnel can begin processing.
Copies of the form can be obtained from any Brooks Instrument location listed above.
2.5 Gas Connections
All models are fitted with the following inlet and outlet connectors as standard: NPT(F), tube
compression fittings, VCR, VCO, DIN or ANSI flanges. Prior to installation, make certain that all
piping is clean and free of obstruction. Install the piping in a manner that permits easy access to
the instrument, should it need to be removed for cleaning or test-bench troubleshooting.