General Description
6 Bruker
BSMS/2 Mainframe Version 001
The following steps may be helpful in troubleshooting such an error:
1. Refer to the installation guide and verify that the installation was done correct-
ly. 
2. Check the power supplies to verify that all the power LED’s (green) are lit. The 
ready LED on the CPU should either be lit or blinking. The ready LED’s on the 
BSMS/2 function boards should be lit. If this is not the case check the power 
supply modules (see Troubleshooting on page 16).
3. Start the BSMS/2 Service Tool ([bsms] in UNIX).
4. In the service tool submenus [6] ‘board functions CPU…’  and [C] ‘board 
functions Keyboard…’ there are functions for debugging the CPU and keyboard, 
respectively.