BSWA 308/309 BSWA Technology Co., Ltd
85
Custom 7
0
0
1
A, Fast, SD
Custom 8
0
0
0
A, Fast, SPL
Custom 9
1
0
0
B, Fast, SPL
Custom 10
2
0
0
C, Fast, SPL
Custom 11
3
0
0
Z, Fast, SPL
Custom 12
0
0
2
A, Fast*, SEL
Custom 13
0
0
3
A, Fast*, E
Custom 14
2
0
6
C, Fast*, Peak
CUSp1_?: Query Custom Measure Setting
Instruction
P1
P2
Explanation
CUS
p1: Group
1~14
Query parameter: ?
ASCII
C
U
S
1
?
Hex
43H
55H
53H
31H
3FH
Byte
1
1
1
1~2
1
Return
Return custom measure setting
Example: query custom measure settings of group 12.
02 01 43 43 55 53 31 32 20 3F 03 1A 0D 0A
Return: the setting of group 12 is A-weighting, Fast, E.
02 01 41 31 32 2C 30 2C 30 2C 30 33 03 6D 0D 0A
TISp1_p2_p3_p4_p5: Set Timer
Instruction
P1
P2
P3
P4
P5
Explanatio
n
TIS
P1:
Switch;
0=OFF;
1=ON;
Default: 0
p2: Start
Day;
0=Ignore;
1~31=
1~31 day
form
today;
Default: 0
p3: Start
hour;
0~23=
0~23h;
Default:
12
p4: Start
minute;
0~59=
0~59m;
Default: 0
P5:
Repeat
period;
1~59=
1~59m;
60~83=
1~24h;
Default: 1
ASCII
T
I
S
0
0
12
0
1
Hex
54H
49H
53H
30H
30H
31H, 32H
30H
31H
Byte
1
1
1
1
1
1~2
1~2
1~2
Return
ACK / NAK