19
CALIBRATION? - Calibration, Cont.
CAL WT 1
- FIRST CALIBRATION WEIGHT
The display will show
CAL WT 1=0
.
Two Point Calibration:
This is the first of two calibration weights. This weight could be ZERO (NO LOAD) or the
TEST WEIGHTS / TEST LOAD.
Three Point Calibration:
This is the first of three calibration weights. This weight could be ZERO (NO LOAD), the
TEST WEIGHTS / TEST LOAD or the MID-POINT WEIGHT .
•
If the first calibration weight is to be ZERO (NO LOAD), press the ENTER key.
• If the first calibration weight is to be the TEST WEIGHTS / TEST LOAD or MID-POINT
WEIGHT, use the numeric keys to input the value of the calibrated test weights.
• Place the weights on the scale platform, then press the ENTER key.
• Starting at the left and proceeding right, a series of dashes will appear on the display.
• The dashes will stay on the display momentarily, then disappear, after which the display
will show:
CAL WT 2
.
CAL WT 2
- SECOND CALIBRATION WEIGHT
The display will show
CAL WT 2=0
.
Two Point Calibration:
This is the second of two calibration weights. This weight could be ZERO (NO LOAD) or
the TEST WEIGHTS / TEST LOAD.
Three Point Calibration:
This is the second of three calibration weights. This weight could be ZERO (NO LOAD),
the TEST WEIGHTS / TEST LOAD or the MID-POINT WEIGHT .
•
If the second calibration weight is to be ZERO (NO LOAD), press the
ENTER
key.
• If the second calibration weight is to be the TEST WEIGHTS / TEST LOAD or MID-
POINT WEIGHT, use the numeric keys to input the value of the calibrated test weights.
• Place the weights on the scale platform, then press the ENTER key.
•
Starting at the left and proceeding right, a series of dashes will appear on the display.
• The dashes will stay on the display momentarily, then disappear, after which the display
will show: CAL WT 3.
CAL WT 3
- LAST CALIBRATION WEIGHT
The display will show CAL WT 3=0.
Two Point Calibration:
This weight is not used. Press the ENTER key to skip and advance to
SERIAL?
.
Three Point Calibration:
This is the last of three calibration weights. This weight could be ZERO (NO LOAD), the
TEST WEIGHTS / TEST LOAD or the MID-POINT WEIGHT .
• If using 3 point calibration and the last calibration weight is to be ZERO (NO LOAD),
press the ENTER key.
• If using 3 point calibration and the last calibration weight is to be the TEST WEIGHTS /
TEST LOAD or MID-POINT WEIGHT, use the numeric keys to input the value of the
calibrated test weights.
•
Place the weights on the scale platform, then press the
ENTER
key.
• Starting at the left and proceeding right, a series of dashes will appear on the display.
•
The dashes will stay on the display momentarily, then disappear, after which the display
will show:
SERIAL?
.