EasyManua.ls Logo

Carel Macroplus - Page 31

Carel Macroplus
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12
Example: If the value is 200 mV ( 0.2 V), the probe signal corresponds to 20 °C/%RH;
0mVd.c. corresponds to 0 ºC.
Pressure probes: In case of extreme out of range reading errors, check, first of all, if the
corresponding analogue inputs of these probes are set to accept current signal - 4÷20 mA
(see pages 8,9). Check, also, that the full-scales set in the software correspond to the probe
full scale. Measuring with a volt-meter the voltage between the terminals Bn and M, you can
calculate the current signal coming from the probe, as the analogue input has an impedance of
50 ohms (I=V/R)
The pressure value "Ps" that the probe transmits can be calculated as follows (FS=Full Scale):
Ps=(Vmis : 50 - 0,004) x (FSmax - FSmin) : 0,016 + FSmin
Example: Probe type: FSmin. -0,5 Bar/FSmax. 7 Bar; Voltage: Vmis= 0,5 Vd.c.
Therefore, the pressure value is:
Ps=(0,5:50 - 0,004) X (7 - (-0,5)) : 0,016 + (-0,5)=2,3Bar
Make sure that the capillary of the probe is not obstructed.
NTC Probes : The probe signal is an 'ohm' value depending on temperature.
The table below shows some resistance values at different temperatures. Disconnect the
probe from the interface and measure its resistance: The corresponding temperature value
can be checked by using the following table.
°C KOhm °C KOhm °C KOhm
-20 67,71 0 27,28 20 12,09
-15 53,39 5 22,05 25 10,00
-10 42,25 17 17,96 30 8,31
-5 33,89 15 14,68 35 6,94
Faulty eeprom alarm:
a. Contact the nearest Service Centre.
Macroplus turns on and off repeatedly (watch dog) or it activates some (digital and/or analogue)
outputs randomly.
Check
a. the connections between interfaces and controller (see pages 2 and 4);
b. power cables: It is recommended that they are sited as far from the interfaces'
microprocessor and the control board as possible;
c. the number of incompatibility errors between the physical and logical outputs of the program
(see operation guide). The software, in fact, might transmit a command that is not carried out
due to a software error or to any interference. Check the outputs to be sure of the correct
functioning of the entire system.