2
8
¾ Sealing Mode :: ( G/N KEY – )
STABLE CONDITION
KEY FUNCTION DISPLAY VAL. DESCRIPTION
Increase or
decrease
set value
S01 =
11
~
99
21: Judge if the weight is changed
Within 1 division for 1 sec
45: Judge if the weight is changed
Within 2 division for 5 sec
99: Judrge if the weight is changed
Within 4.5division for 9 sec
AUTOMATIC ZERO TRACKING ( 2% of MAX. CAPACITY)
KEY FUNCTION DISPLAY VAL. DESCRIPTION
Increase or
decrease
set value
S02 =
11
:
99
21: Judge if the weight is changed
Within 1 division for 1 sec
45: Judge if the weight is changed
Within 2 division for 5 sec
99: Judrge if the weight is changed
Within 4.5 division for 9 sec
Kg/lb CONVERSION
KEY FUNCTION DISPLAY VAL. DESCRIPTION
set value
↓↑
ENTER Go to Next
S03 =
0
1
Not used kg/lb conversion
Use kg/lb conversion