30
¾ Zero Calibration :: ( PRINT KEY - )
ZERO CALIBRATION
KEY FUNCTION DISPLAY DESCRIPTION
ULoAd
Remove weight from the platter
and Press ENTER KEY
---------
Performing Zero Calibration
ENTER
After setting
the A/D value,
go to initial
CALMOd
Good
Zero Calibration is finished
Note 1. Only Zero Calibration is performed When the zero value exceeds the initial
Zero range in this mode.
¾ Engineer Mode :: (PRINT LONG KEY - )
Engineer Mode
KEY FUNCTION DISPLAY VAL. DESCRIPTION
0
ENTER
UNIT
EXIT Mode
FAC 0
1
It’s not item for user to set the value
Please exit to press the ENTER key.
¾ Exit Mode :: ( UNIT KEY - )
After calibration mode, go to weighing mode.
Sealing Method
After calibration, you have to seal as follows.
1. Fasten the CAL Switch Bolt.
2. Connect the sealing wire as figure.
3. Press the sealing wax as figure.