29Method - Create, Edit and Delete
5. Select the “Parameters” tab.
5.1. Once in the system parameters, determine if the stage will be a Set Time or Constant Weight analysis.
NOTE
If creating multiple stages, only the nal stage can be a Constant Weight stage. All previous stages must
use Set Time.
5.2. Set Time - The method will run for a preset length of time, regardless of sample size or nal weight.
5.2.1. Set the Mag Power and IR power . The power should be set at 0%, 25%, 50%, 75%, or 100%
5.2.2. Set the run time for each analysis.
5.2.3. Set the Maximum temperature for the sample type. Some category parameters contain a con-