Instruction Manual CertoClav Vacuum Pro 45 l Model STE-45-T
19
7 Sterilisation Programs
7.1 Default settings
Program
Vacuu
m
phases
Sterilisation
time [min]
Sterilisation
temperature
[°C]
Sterilisation
pressure
[kPa]
Vacuum
retention time
[min]
Drying time
[min]
134°C SOLID
1 4 134 210 - 5
VACUUM TEST
1 - - -75 10 15
121°C WRAPPED
3 20 121 110 - 5
134°C WRAPPED
3 4 134 210 - 5
B&D TEST
3 3.5 134 210 - -
HELIX TEST
3 3.5 134 210 - -
134°C PRION
3
-
134 -75 15 -
121°C LIQUID
- 30 121 110 - -
The SOLID program should only be used for unwrapped solids. Hollow bodies of type
A and B or porous materials cannot be sterilised with this program.
The VACUUM TEST program performs a pressure leak test.
The WRAPPED programs are suitable for the sterilisation of porous material, hollow
bodies of type A and B, packaged waste and instruments. Both programs do not differ
in end result. If the load is temperature sensitive and cannot therefore be sterilised at
134°C, please select the 121°C program.
The B&D TEST or HELIX TEST program tests hollow bodies of type A. The programs
are almost identical. The only difference involves the fact that the B&D TEST program
is a B&D test package, while the HELIX TEST is used as a test device. (PCD: Process
Challenge Device).
The 134°C PRION program is specially designed for the sterilisation of prions on all
kinds of surface.