Instruction Manual CertoClav Vacuum Pro 45 l — Model STE-45-T
58
E5
Pressure relief
error
The pressure could not be
released from the chamber
Contact your dealer.
E6 Door sensor error
The door is open during
sterilisation
Contact your dealer.
E9
Temperature
The temperature could not be
Contact your dealer.
E11
Steam generator
The steam generator could
Contact your dealer.
E12 Chamber heating
The jacket heating could not
Contact your dealer.
E13
Vacuum pump or
leak tightness
The generation of a vacuum
during a cycle has failed
Clean the door seal and restart
the process. Contact your dealer
if the error persists.
E20
Program
terminated
The user has terminated the
program manually during a
E22
Vacuum pump or
leak tightness
Vacuum test has failed
Clean the door seal and restart
the process. Contact your dealer
if the error persists.
E23
Vacuum test was
terminated
unexpectedly
The temperature in the
chamber has become too hot
during the program cycle
Deactivate the “Keep warm” or
“Retain temperature” function in
the sterilisation settings and wait
until the chamber has cooled
E24
The pressure is
less than 30 kPa at
an temperature of
The temperature-pressure
ratio does not correspond to
that of saturated steam.
Contact your dealer.
E25 Door locking failed
The door could not be locked.
The door is possibly set too
Contact your dealer.
E26
Door opening
Contact your dealer.
E27
The vacuum test
cannot be started
The temperature in the
chamber is too hot to start the
vacuum test
Deactivate the “Keep warm” or
“Retain temperature” function in
the sterilisation settings and wait
until the chamber has cooled
down.