What to do if liquid is coming out of the dosing head of my CHART ULTRADOSER 2K Medical Equipment, even when the valve is closed?
- AAshlee DavisAug 9, 2025
If liquid continues to flow from the dosing head of your CHART Medical Equipment despite the valve being closed, the LN supply pressure might be too high; try reducing it. Alternatively, the valve seat could be contaminated with ice or particles, requiring the unit to be drained of LN and the dosing valve assembly cleaned. Also, ensure the air supplied to the solenoid is not reversed; if it is, reverse the LN lines that supply and vent to the solenoid.