Calibration
From [Setup], select [Calibration] to set up the dosing parameters. Determine the desired Dose Mode:
Auto or Fixed. Auto Dose Mode compensates for any changes in line speed by auto-adjusting the
distance delay and dose duration. Fixed Dose Mode does not compensate for any changes in line speed.
The distance delay and the dose duration will remain unchanged.
IMPORTANT: Proper set up of the [Calibration] screen is essential to the successful operation of the
UltraDoser 2K system. If after reading this section you are still unsure, please contact Chart’s service
group at +1 408-371-9313.
Auto Dose Mode
Enter Setup Parameters for Auto Dose Mode:
1. Measure and enter the distance between two containers, centerline to centerline in the
‘Container CL to CL’ field. See image below for additional information in determining this
value.