74 DF II Series User Manual
DFE II DFS II DFS II-R DFS II-R-SLC DFS II-R-STS
Accuracy: % of full scale +0.25% +0.1% +0.1% +0.25%
1
0.3%
Data Sampling Rate (KHz): 7 10 10 10 10
Peak Capture Rate (Hz): 7000 10,000 10,000 10,000 10,000
Display Update Rate (Hz): 10 10 10 10
Display Resolution: 1000:1 10,000:1 10,000:1 10,000:1 10,000:1
Tare Capacity (% Full Scale): 110% 110% 110% 110% 110%
Overload Protection (% Full Scale): 150% 150% 150% 150% 150%
Display Type: Color Color Color Color Color
Memory (No. of Saved Results): 20 100 100 100 100
Note
1
: SLC-02000, SLC-05000 and SLC-10000 capacities used with the DFS II-R-ND have a calibration accuracy of +0.5% of full scale.
Common Specications to all DF Models
Temperature Rating
Operating Temperature: 40
O
to 100
O
F (4
O
to 38
O
C)
Storage Temperature: 0
O
to 130
O
F (-17
O
to 54
O
C)
Temperature Stability: +0.015% per
O
F
Relative Humidity: 20% to 85% non-condensing
Conformity: BS EN 61010-1: 2010 Safety Requirements for Electrical Equipment, BS EN 61326--1: 2006 Electrical
Equipment for Measurement, Control and Laboratory use: General Requirements BS EN 61326--2-3: 2006 Electrical
Equipment for Measurement, Control and Laboratory use: Transducers with integrated or remote signal conditioning,
RoHS and WEEE
Power: Battery (Nickel Metal Hydride) or direct AC 120/230Vac
Battery Life: With Dimming ON and Bluetooth OFF: 20 Hours
With dimming OFF and Bluetooth OFF: 16 Hours
With dimming ON and Bluetooth ON: 16 Hours
With dimming OFF and Bluetooth ON: 12 Hours
Software: Use with TCD WEDGE application software (Nexygen DF II)
Instrument Weight: 1.5 lbs (0.7 kg)
Shipping Weight: 4 lbs (2 kg)
Warranty Period: 1 year
FEATURES & FUNCTIONS
Operating Modes DFE II DFS II DFS II-R DFS II-R-ND
Normal Std Std Std Std
Peak (Tension & Compression) Std Std Std Std
Load Limits Std Std Std Std
Load Average Std Std Std Std
Break Detection - Std Std Std
Contact Closure - Std Std Std
Peak (Torque) - - - Std
Tolerances (Pass/Fail) Std Std Std Std
Outputs
Analog Std Std Std Std
Digital RS232 Std Std Std Std
USB Std Std Std Std
Bluetooth® - Std Std Std
Mitutoyo Std Std Std Std
Output Filters
Peak Std Std Std Std
Display Std Std Std Std
Statistics
Mean Std Std Std Std
Max/Min Std Std Std Std
COV Std Std Std Std
Standard Deviation Std Std Std Std
% Difference Std Std Std Std
Password Std Std Std Std
Auto Dimming Std Std Std Std
Multiple Languages Std Std Std Std
Graphing Std Std Std
SPECIFICATIONS

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Chatillon DFE I and is the answer not in the manual?

Chatillon DFE I Specifications

General IconGeneral
BrandChatillon
ModelDFE I
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals