Electrical Safety Analyzer 19032-P Quick Start Guide
1
1. Introduction
1.1 An Overview of Product
Automatic withstand / insulation / grounding testers of the instrument are
designed for automatic withstand, insulation resistance, grounding
resistance, short/open circuit detection and dynamic leakage current test of
electromechanical and electronic equipments.
The aspect of withstand voltage testing, the output power is AC: 500VA(5kV,
100mA), DC: 150VA(6kV, 25mA). Therefore, it is for withstand test of
electronic and electromechanical and component.
Testing aspect of insulation resistance, the measurement range is 0.1MΩ
50GΩ and test voltage range is 50V 1000V can be set arbitrary.
Testing aspect of grounding resistance, the grounding resistance range can
be measured is 10 150mΩ, under 10A can up to 510mΩ. The output
test current range is 3 40A can be set arbitrary.
Testing aspect of dynamic leakage current, the measurement range is
0.01mA 50.0mA (rms). The output test voltage range is 90V 280VAC.
The test rule matches to IEC950, UL544, UL2601 etc.
In the testing aspect of short/open circuit detection, please test if
capacitance is short or open before testing high voltage. Please make
sure the DUT good contact then processes high voltage test.
All of setting status, time, current, voltage, resistance value, memory
number etc are list on the display, it is unnecessary to remember any
parameter status which be set.
The tester equipped with Good and No Good judgment machinery and
signal output of testing result and remote control. It is also for GPIB
interface, SCANNING interface, RS232 interface of automatic test system.
The above equipments makes high efficient and accurate test.