Electrical Safety Analyzer 19032-P Quick Start Guide
48
4.14 Output Signal
The analyzer includes LED and buzzer two kinds of indication signals.
The rear panel of analyzer has the following output signals.
UNDER TEST: When the analyzer is under test, the output terminal will
short circuit. Be able to use this short circuit condition to
control external signal. The junction specification 125VAC
current is lower than 1A.
PASS: When the analyzer judge DUT is good, the output terminal will short
circuit. Be able to use this short circuit condition to control
external signal. The junction specification 125VAC current is
lower than 1A. Operating time is from DUT judged as pass to be
stopped or restart.
FAIL: When the analyzer judge DUT as no good, the output terminal will
short circuit. Be able to use this short circuit condition to control
external signal. The junction specification 125VAC current is lower
than 1A. Operating time is from DUT judged as fail to be stopped or
restart.
EOS: When the analyzer performs the test in test step, the output terminal
will be short-circuited. Be able to use this short circuit condition to
control external signal. The junction specification 125VAC current is
lower than 1A.