Chroma 19032-P - Start Wait Function

Chroma 19032-P
63 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Electrical Safety Analyzer 19032-P Quick Start Guide
TEST
MODE SOURCE
L
IMIT
R
ES.
O
FFSET
29
01 AC 1.000
k
V
0
.
5
03
m
A
G
et Cs
P
AGE UP
P
AGE DOWN
SCANNER-1
1 2 3 4 5 6 7 8
AC H
TEST TIME: 0.0s
R
e
mote Lock
o
ffset
E
rror
Figure 4-5
4.6.4 Start Wait Function
1. It sets Start Wait time, for example: the setting is 3 seconds.
2. According to paragraph 4.8.2 for setting various parameters of GB
MODE. Take for example: CURRENT set as 25.00A, HIGH LIMIT:
100mΩ, TEST TIME: 3.0sec. TEST screen is as Figure 4-6 shown.
TEST
MODE SOURCE LIMIT RES. OFFSET
01 GB 25.00
A
1
00.0
m
Ω
et Cs
P
AGE UP
P
AGE DOWN
SCANNER-1
1 2 3 4 5 6 7 8
GB
emote Lock
o
ffset
E
rror
Figure 4-6