Electrical Safety Analyzer 19032-P Quick Start Guide
2
2. Specification (18°C ∼ 28°C RH ≤ 70%)
AC/DC Withstanding Test
Output Voltage
AC: 0.05-5.0 kV, steps 0.001kV, DC: 0.05-6.0
kV, steps 0.001kV.
Load Regulation
≤ (2% of setting + 0.1% of full scale), Rated
load
Voltage Accuracy ± (2% of setting + 0.1% of full scale)
Cutoff Current
(Note 1)
AC: 0.1mA ~ 100mA, DC: 0.01mA ~ 25mA
0.1uA DC resolution
V-display Accuracy
± (1% of reading + 0.1% of full scale), 2V
resolution
Leakage Current
Meter
AC current:
3mA range: 0.001mA - 2.999mA, 0.001mA
resolution
30mA range: 0.01mA - 29.99mA, 0.01mA
resolution
100mA range: 0.1mA - 100.0mA, 0.1mA
resolution
Measurement Accuracy: ± (2% of reading +
0.5% of range)
DC current:
300uA: 0.1uA– 299.9uA, 0.1uA resolution
3mA range: 0.001mA – 2.999mA, 0.001mA
resolution
25mA range: 0.01mA – 25.00mA 0.01mA
resolution
Measurement Accuracy: ± (2% of reading +
0.5% of range)
Output Waveform
50Hz, 60Hz ± 0.1%, sine wave. sinewave,
Crest Factor:1.3~1.5
Test Time(Note 2) 0.3 ~ 999 Sec. Continue
Ramp Time 0 ~ 999 Sec. off
Fall Time 0 ~ 999 Sec. off
DWELL Time 0 ~ 999 Sec. Off (WDC only)
Maximum Short
Current
Up to 4kV 200mAac only
Flashover (ARC) Detection (SPC) (Note 3)