Electrical Safety Analyzer 19032-P Quick Start Guide
4.2 Rear Panel
(1) (2) (3) (4) (5) (6) (7) (8) (9) (10) (11) (12)
17
Figure 4-2
(1) HV2: Reference terminal of high voltage output terminal (HV1) is low
electric potential terminal.
(2) HV1: High electric potential terminal of high voltage output.
(3) Rear Panel GB Output Terminal: This terminal Floating status is
selectable, that is open-circuited with front panel GB terminal.
(4) USB: USB terminal
(5) SCAN Interface: This interface can connect with 9030A Scanning Box
(Option).
(6) GPIB Interface(Option): This socket is for optional GIPB interface
(IEEE-488-1978).
(7) RS232 Interface: This socket is for RS232 interface of the instrument.
GPIB and RS232 interface can’t be used simultaneously.
(8) Plug in SCANNER Insert Hole (Option):
Option (6000-01): 5 points of
HV output & 3 (UL approval)
sets of grounding terminals.
Option (6000-02): 3 points of
HV output & 5 (UL approval)
sets of grounding terminals.