Electrical Safety Analyzer 19032-P Quick Start Guide
3
Detection Current
AC: 1mA – 20mA, DC: 1mA – 10mA, resolution
0.1mA
Min. pulse width 40us 20us 10us 4us Approx.
TwinportTM function (SPC)
Functions
WV and GB test can be performed at the same
time. (ON/OFF)
Insulation Resistance Test (Note 4)
Test Voltage DC: 0.05kV ~ 1kV, Constant Voltage
V-display Accuracy ± (2% of reading + 0.5% of full scale)
Resistance Range
0.1MΩ ∼ 50GΩ
Measuring
Accuracy
≥ 500V:
1 ∼ 1000MΩ: ± (5% of reading + 0.5% of full
scale)
1001 ∼ 9999MΩ: ±(10% of reading + 0.5% of
full scale)
10GΩ ∼ 50GΩ: ±(15% of reading + 1% of full
scale)
< 500V:
0.1 ∼ 1000MΩ: ±(10% of reading + 0.5% of full
scale)
< 100V:
0.1 ∼ 1000MΩ: ±(15% of reading + 0.5% of full
scale)
Test Time(Note 2) 0.3 ~ 999 Sec. Continue
Ground Bond Test (Note 5)
Output Current 3.00 ~ 40.00Aac. Constant Current, 0.01A step
Current Accuracy ± (2% of setting + 0.1% of full scale)
Output Waveform 50Hz, 60Hz ± 0.1%, sine wave
Current Meter
0.00 ∼ 40.00A
Meter Accuracy ± (1% of reading + 0.17% of full scale)
Resistance Range
10.0 ~ 510.0mΩ (with offset value)
Resistance
Accuracy
± (2% of reading + 0.1% of full scale) at ≥ 10A
± (2% of reading + 1% of full scale) at
<10A(<210mΩ)
± (3% of reading + 2% of full scale) at
<10A(>210mΩ)