24. Descriptions of ground bond lead wiring
The maximum output current of this unit is 40amp AC, a no good
connection will cause temperature rising and ground bond output
terminal may be burned down. Please follow the below steps to make
good wiring connection.
• Using box spanner to tighten up test cable of DRIVE+ and DRIVE-.
25. Keep test cable away from the panel
Please keep the high voltage cable or the DUT away from the panel at
least 30 cm during operation to avoid the display interference caused
by high-voltage discharge.
26. Notices for connecting automated device
• The grounding system of the device and the automated station
should be connected together.
• Add anti-interference iron core to the high voltage cable and the 2
ends (device output and DUT) of RTN/LOW test cable with winding
at least 1 circle.
• The high voltage and RTN/LOW test cable must be separated from
the control cable.
• The high voltage and RTN/LOW test cable must keep proper
distance from the analyzer panel.