Electrical Safety Analyzer 19032-P Quick Start Guide
Source, please set it as ON.
After Fail CONTINUE /
RESTART /
STOP
RESTART (1) When set as CONTINUE, and
any one among STEPs judged
as No Good. It will continue
until all STEPs are tested.
(2) When set as START, and any
one among STEPs judged as
No Good press START to
restart directly.
(3) When set as STOP, and any
one among STEPs judged as
No Good. It is necessary to
press STOP then can restart
test by pressing START.
AC OFFSET 0 ~ 2.5mA 0.10mA (1) When Offset value is higher
than AC OFFSET value,
Current reading = Current real
measurement value – Offset
value.
(2) When Offset value is lower
than AC OFFSET value,
Current value =
22
(Real measurement value) ( )Offset
LC OFFSET 0 ~ 2.5mA 0.00mA (1) When Offset value is higher
than LC OFFSET value,
Current reading = Current real
measurement value – Offset
value.
(2) When Offset value is lower
than LC OFFSET value,
Current value =
22
(Real measurement value) ( )Offset
LC OFFS
GET
ON/OFF ON (1) When the setting is ON, LC
Mode will be included as
processing OFFSET GET.
(2) When the setting is OFF, LC
Mode won’t be included as
processing OFFSET GET.