Electrical Safety Analyzer 19032-P Quick Start Guide
36
Ground Resistance Test Mode (GB)
CURRENT: It sets ground resistance test needed current.
Notice: Because the high limit of multiplying test current by resistance can’t
higher than 6.3V. High limit of resistance will auto modify to
adaptable value when it isn’t correspondence with the above
conditions.
HIGH LIMIT: It sets ground resistance judgment high limit value. The high
limit value is 510mΩ or 6.3V/CURRENT.
LOW LIMIT: It sets ground resistance judgment low limit value, the range is
from 0 to high limit of resistance. Input 0 means OFF.
TEST TIME: It sets test needed time. Input 0 means continuous test.
TWIN PORT: It selects twin port, can select ON / OFF. When set as ON,
and next STEP is AC/DC or IR, the two steps can operate
simultaneously. The highest AC rated current when twin port
can’t over 5kV 50mA and GB current can’t over 20A, or it may
cause output voltage, current distortion.
CHNL (H-L): It sets scan test point (please set with optional device, for
example 6000-01).
Withstand Voltage Test Mode (AC)
VOLTAGE: It sets withstand voltage test needed voltage
FREQ.: It sets AC withstand test signal frequency, input 0 to indicate
DEFAULT that is to follow the frequency which set in section 4.6 for
testing.
HIGH LIMIT: It sets high limit value of leakage current.
LOW LIMIT: It sets low limit value of leakage current. The range is lower
than high limit value of leakage current or OFF.
ARC LIMIT: It sets high limit value of arc.
ARC FILTER: It selects frequency range of detection arc. There are four
frequency ranges of 323 kHz/350 kHz/3100 kHz/3230
kHz can be selected.
TEST TIME: It sets test needed time. It inputs 0 means continuous test.
RAMP TIME: The needed time which rises to setting voltage. It inputs 0
means OFF.
FALL TIME: The needed time which falls from setting voltage value to
zero, 0 means OFF.
CHNL (H-L): It sets GB-Floating test selection point.
(1) When set CHANNEL 3 to H (high):
(a) Start test: HV1 terminal on front panel and that on rear panel are
short-circuited with high voltage output. HV2 terminal on front
panel and that on rear panel are short-circuited with low voltage
terminal. DRIVE and SENSE terminals on front panel and that on

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Chroma 19032-P and is the answer not in the manual?

Chroma 19032-P Specifications

General IconGeneral
BrandChroma
Model19032-P
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals