High Power DC Electronic Load 63200 Series Operation & Programming Manual
7.2.8 POWer Sub-system
POWer
:L1
:L2
:RISE
:FALL
POWer:L1/L2
Description Set the Static Load Current of constant power mode.
Syntax POWer:L1 <NRf+>[suffix]
POWer:L2 <NRf+>[suffix]
Parameters For valid value range refer to respective specification.
Example POW:L1 20 Set Constant Power = 20W for Static Load L1.
POW:L2 10 Set Constant Power = 10W for Static Load L2.
POW:L1 MAX Set Constant Power = max. value for Static Load
L1.
POW:L2 MIN Set Constant Power = min. value for Static Load
L2.
Query Syntax POWer:L1?
POWer:L2?
POWer:L1? MAX
POWer:L2? MIN
Return Parameters <NR2> [Unit=Watt]
Query Example POW:L1? Return the set current value of the Static Load
L1.
Return Example 3.12
POWer:RISE/FALL
Description Set current slew rate of constant power static mode.
Syntax POWer:RISE <NRf+> [suffix]
POWer:FALL<NRf+> [suffix]
Parameters For valid value range refer to respective specification.
Example POW:RISE 2.5 Set rise slew rate to 2.5A/µS for static load.
POW:FALL 1A/µS Set fall slew rate to 1A/µS for static load.
Query Syntax POWer:RISE?
POWer:FALL?
POWer:RISE? MAX
POWer:FALL? MIN
Return Parameters <NR2> [Unit=A/µS]
Query Example POW:RISE? Return the rise slew rate of static load.
Return Example 2.5
7-20