GENERAL DESCRIPTION
22
Table 2-4: ZP Load Capacity and Electrical Requirements
Cascade -70° C to 190° C
Live Load Capacity
Watts
Electrical Power Requirements
Full Load Amps
Model
-18° C
-34° C
-40°C
-54°C
-68°C
208/230V, 1Phase
208/230V, 3Phase
460V, 3Phase
ZP-8-2-2-H/AC
-
-
1200
900
600
49
36
-
ZP-8-3.5-3.5-SC/AC
-
-
1700
1200
750
-
48
29
ZP-16-2-2-H/AC
-
-
1800
1200
600
49
36
-
ZP-16-3.5-3.5-SC/AC
-
-
2000
1600
1000
-
48
29
ZP-16-6-6-SC/WC
-
-
3000
2400
1500
-
65
29
ZP-32-2-2-H/AC
-
-
1500
1100
600
49
36
-
ZP-32-3.5-3.5-SC/AC
-
-
2000
1600
1000
-
48
29
ZP-32-6-6-SC/WC
-
-
3000
2400
1500
-
65
29
ZP-32-10-10-SC/WC
-
-
4500
3500
1700
-
-
45
ZP-32-15-15-SC/WC
-
-
5500
4500
2500
-
-
53
ZP-44-3.5-3.5-SC/AC
-
-
1700
1300
700
-
48
29
ZP-44-6-6-SC/WC
-
-
2700
2100
1200
-
65
29
ZP-44-10-10-SC/WC
-
-
4500
3500
2000
-
-
45
ZP-44-15-15-SC/WC
-
-
5500
4500
3000
-
-
53
ZP-64-3.5-3.5-SC/AC
-
-
1500
1100
500
-
48
29
ZP-64-6-6-SC/WC
-
-
2500
2100
1000
-
65
29
ZP-64-10-10-SC/WC
-
-
4500
3500
2000
-
-
45
ZP-64-15-15-SC/WC
-
-
5500
4500
3000
-
-
53
ZP-80-3.5-3.5-SC/WC
-
-
1500
1100
500
-
48
29
ZP-80-6-6-SC/WC
-
-
2500
2100
1000
-
65
29
ZP-80-10-10-SC/WC
-
-
4500
3500
2000
-
-
45
ZP-80-15-15-SC/WC
-
-
5500
4500
3000
-
-
53
ZP-96-3.5-3.5-SC/AC
-
-
1500
1100
500
-
48
29
ZP-96-6-6-SC/WC
-
-
2500
2100
1000
-
65
29
ZP-96-10-10-SC/WC
-
-
4500
3500
2000
-
-
45
ZP-96-15-15-SC/WC
-
-
5500
4500
3000
-
-
53

Table of Contents