X1 Series Dental Unit -Operating Manual
Add: Building 6-1,Nansha Industrial Village Jiangli North Rd,Danzao Town, Nanhai District,Foshan ,Guangdong,China
www.cingol.com
the balance arm,and fix with screw.
F)Connect power supply
Put the plug into the power socket,which must have the wire connected to
ground.the wire should have the copper wire inside less than 1mm make sure
this wire connected well with ground.
3.6 Modulation
Now the device is at state of modulation after different parts are fitted well.
Open the base box,check the air pressure meter of air filtering adjusting valve
whether it reads 0.6mpa,if not, adjust the valve to keep the air pressure reading
0.6mpa.(Method: pull up the rotary knob at the top air pressure valve,turn
clockwise for increasing pressure, turn anticlockwise for reducing pressure.