7 Malfunctions
Some common fault is following table:
The mebrane of the pile-up valve has been
damaged
The water valve switching on by air has
been damaged
The O-ring or the spring of the water button
has been damaged
Change a new O-ring or a spring.
The tumbler filler
can’t be shut off.
Remove and clean the valve core.
The flusher can’t
be shut off.
Remove and clean the valve core.
The tumbler filler
can’t open.
The solenoid valve has been damaged
Change a new solenoid valve.
The circuit board in the water unit has been
damaged
Change a new circuit board
The solenoid valve has been damaged
Change a new solenoid valve.
The circuit board in the water unit has been
damaged
Change a new circuit board
The heater has been damaged
The circuit board in the water unit has been
damaged
Change a new circuit board
The saliva ejector
is not enough
suction
Water pressure is not enough.
The filter has been blocked by dirt
Open and clean the filter
The spray
aspirator is not
enough suction
Air pressure is not enough.
Increase air pressure by adjusting
the reducing valve
The filter has been blocked by dirt
Open and clean the filter
The patient chair
is malfunction
The motor has been damaged
The circuit board controlling
motors has been damaged
Change a new circuit board
Find out the cause and change a
new fuse.
The circuit board in the operating light has
been damaged
Change a new circuit board
Fuse of the operating light
damaged
Find out the cause and change a
new fuse.