X1 Series Dental Unit -Operating Manual
Add: Building 6-1,Nansha Industrial Village Jiangli North Rd,Danzao Town, Nanhai District,Foshan ,Guangdong,China
www.cingol.com
5 Cleaning and maintenance by the practice team
To maintain the value and safe functioning of your dental unit, it is
necessary to have it cared for, cleaned and disinfected by the practice
team regularly. This will minimize the risk of contamination for patients
and users and ensure proper functioning.
Remove any dirt and disinfectant residues using a mild commercial
cleaning agent.
The time intervals, operational approach and cleaning agent refer to
the following table.
Care and cleaning intervals
The time intervals specified for disinfection, care and cleaning are
reference values.
Please adapt the time intervals to suit your personal method of working
and your national requirements.
Medicaments chemically react with the surface of the unit.
Due to their high concentrations and the substances they contain, many
medicaments can dissolve, etch, bleach or discolor surfaces.
Clean any medicament residues off of the unit immediately with a
moist cloth!
Liquids can enter the unit during cleaning or disinfection
Electrical components of the dental unit can be destroyed by liquids.
Do not spray any liquids into the unit.
To clean near openings, first spray the liquid onto a cleaning cloth.
Then wipe over the unit with the cleaning cloth.
Disinfectants can dissolve dyes in cleaning clothes.
The outer surface of the unit may then be discolored by the dye.
Do not clean or disinfect the unit with colored cleaning clothes.