X1 Series Dental Unit -Operating Manual
Add: Building 6-1,Nansha Industrial Village Jiangli North Rd,Danzao Town, Nanhai District,Foshan ,Guangdong,China
www.cingol.com
8
3 Unit description
3.1
Standards/approvals
The X1 dental unit complies with the following standards, among
others:
IEC 60364-7-710
IEC 60601-1:2005+A1:2012
IEC 60601-1:2006+A1:2013
IEC80601-2-60
ISO 7494-1:2018
ISO 7494-2:2003
ISO 9680:2007
ISO 13485:2016
This product bears the CE mark in accordance with the provisions of
Council Directive 93/42/EEC of 14 June 1993 concerning medical
devices.
3.2
Technical data
Model designation: X1
Input max power: 600VA
Power connection: 230 V AC ± 10%
50/60 Hz
Rated current: 3A at 230 V
Fuse for the unit: Ø5 250V AC 6A slowblow
Protection class: Class I equipment
Degree of protection
against electrical shock: Type B applied parts
Degree of protection Ordinary equipment (without protection
against ingress of water: against ingress of water)
The foot control has an IP X1 degree of
protection against liquids (drip-proof).
Mode of operation: Continuous operation with intermittent
loading corresponding to the dental mode
of working.
Permanently connected unit.