X1 Series Dental Unit -Operating Manual
Add: Building 6-1,Nansha Industrial Village Jiangli North Rd,Danzao Town, Nanhai District,Foshan ,Guangdong,China
www.cingol.com
2.5
Maintenance and repair
Authorized technical As manufacturers of dental units and in the interest of the operational
personnel and spare safety of your unit, we stress the importance of having maintenance
parts and repair of this unit performed only by ourselves or by agencies
expressly authorized by us. Furthermore, failed components must
always be replaced with original spare parts.
Maintenance intervals In order to ensure the operational safety and reliability of your unit and
to avoid damage due to natural wear, you as the unit owner must have
your unit checked regularly by an authorized service engineer.
Furthermore, safety checks must be performed. Please contact your
dental depot to obtain a maintenance offer. For more information,
please refer to “Maintenance by the service engineer”.
2.6
Trouble-free operation
Use of this unit is permissible only if it works properly without
malfunctions. If trouble-free operation cannot be ensured, the unit
must be taken out of service, checked by authorized technicians for
malfunctions and, if necessary, repaired.
2.7
Patient chair
Please observe the maximum load capacity of 185 kg for the patient
chair.
The weight distribution complies with the ISO 7494-1:2018 standard.
The safety test is according to IEC60601-1:2007.
The patient’s arms and legs must be testing on the upholstery of
the chair.
2.8
Ventilation slots
Under no circumstances may the ventilation slots on the unit be
covered, since otherwise the air circulation will be obstructed. This can
cause the unit to overheat.
Do not spray liquids such as disinfectants into the ventilation slots.
This may lead to malfunctions. Use wipe disinfection only in the vicinity
of the ventilation slots.
2.9
Intermittent operation
The motors of the unit are designed for intermittent operation
corresponding to the dental mode of treatment.
Drive motors for patient chair and backrest: max. 10% duty cycle,
cycle duration 300s