The dental unit must not be operated if it has failed to pass the safety
tests!
6 Maintenance by the service engineer
6.1
Inspection and maintenance
In order to ensure the operational safety and reliability of your dental
unit and to avoid damage due to nature wear, annual inspection and
maintenance must be performed on your dental unit. This is done by
an authorized service engineer.
An overview of the inspection and maintenance work performed is
additionally recorded by the service engineer in the “Warranty
Passport”.
6.2
Safety checks
Medical equipment is designed in such a way that the first occurrence
of a fault does not create a hazard to the safety of patients, users or
other persons. It is therefore important to detect such faults before a
second fault occurs, which might then lead to safety hazards.
For that reason it is essential to perform safety tests every 2 years
which aim particularly at detecting electrical faults (e.g. isolation
defects). This is done by an authorized service engineer.