X1 Series Dental Unit -Operating Manual
Add: Building 6-1,Nansha Industrial Village Jiangli North Rd,Danzao Town, Nanhai District,Foshan ,Guangdong,China
www.cingol.com
Disinfect water lines of the
instruments
Clean the water and air filters
Change the cotton wool in the
return-air bottle
Clean the filter of the suction
device
disinfect the water line of the
cuspidor bowl
Disinfect the suction hoses
Cleaning And Maintenance
Items
1. Remove the pure water bottle and fill disinfectant.
Screw up the bottle again. Pull the air switch to
the NO position and pull the water alternative
switch to the Bottle Water position.
2. Removed the instruments from their holders and
held over the cuspidor bowl; step on the water
pedal of the foot control for purging. Press the
water key of the 3-way syringe for purging. The
purging time is for least 30 minute.
3. Purge the water lines again with the pure water.
Open the water/air filter. Remove the filter element and
clean it with cleaning water.
Unscrew the return-air bottle and remove the cotton
wool roll. Insert new cotton wool roll and screw up the
bottle.
Open the filter flask and clean the filter with
disinfectant.
Wipe over surface with a moist cloth (water).
Pour 1 liter of the disinfectant into the cuspidor for
flushing the water line.
Wipe over surface with disinfectant.
Pumping 500 ml disinfectant to flushing the suction
hoses.
1000mg/L available chlorine
1000mg/L available chlorine
1000mg/L available chlorine