- 23 -
14. PROCEDURE OF CALIBRATION
If moisture balance is used as II lass laboratory scale do the
calibration before measurement.
Calibration defines relations between value if annunciator and mass
of standard (calibration weight), which is load of balance and corrections if
it’s meccesary.
To calibrate scale calibration weight of class F
2
is neccesary. It’s
additional equipment for moisture balance.
Type of scale Calibration weight
MB 200 200g
Procedure of calibration
• Press TARE key, when horizontal bars appears on display –
press F.
• following order appears on display
• press TARE – leave calibration
Order to put calibration weight with defined mass on scale
Calibration is automatic after weight is put on scale. Following
order appears on display: