9. CONTROL RELIABILITY OF TEMPERATURE ANNUNCIATORS
Checking if temperature of drying is conformable to temperature shown
on display of moisture balance. To control certified thermometer PT 101 is
used. It’s additional equipment of every moisture balance.
Both terometers measure temperature of air in short distance from
controled sample.
10. PROGRAMMING PARAMETERS OF DRYING
Drying is automatical. In ranges of time settled by user (time of testing
samples) balance compares mass of sample with two previous
measurement and shows moisture without delay. If 3 following
mesurements of sample mass are the same counted value is kept on
digital display, automatic measurement is finished automatically. Automatic
finish can be earlier if time of drying settled by user is shorter (refers to
rnodE 4,5,6,7)
Before drying all values should be programmed. Press key:
After pressing key F display shows order SubSt.