EasyManua.ls Logo

CLARIUS HD3 - Clarius Scanner PAL HD3: B-Mode, Linear Array

Default Icon
142 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Clarius Scanner HD3
revision 9 124
Clarius Scanner PAL HD3: B-Mode, Linear Array
Index Label MI TIS TIB TIC
At
Surface
Below
Surface
At
Surface
Below
Surface
Maximum Index Value 0.349 0.046 0.067 (a)
Index Component Value
0.046 0.046 0.067 0.046
Acoustic
Parameters
p
r,a
at Z
MI
(MPa) 0.989
P(mW) 1.192 1.192 #
P
1x1
(mW) 1.192 1.192
z
s
(cm) 1.953
z
b
(cm) 1.953
Z
MI
(cm) 1.953
Z
pii,a
(cm) 1.953
f
awf
(MHz) 8.045 8.045 8.045 #
Other Information prr (Hz) 5376
srr (Hz) 24.0
n
pps
2
I
pa,a
at Z
pii,a
(W/cm
2
)
47.092
I
spta,a
at Z
pii,a
or Z
sii,a
(mW/cm
2
)
1.533
I
spta,a
at Z
pii
or Z
sii
(mW/cm
2
)
4.527
p
r
at Z
pii
(MPa) 1.699
Operating Control
Conditions
Control 1 
Control 2
Control 3
Control 4
Control 5
Control 6
Control 7
(a) This scanner is not intended for transcranial or neonatal cephalic use.
# Not reported for this operating condition; the maximum index value is not reported for the reasons shown in the Maximum
Index Value fields.
Control 1: Exam Type: Vascular; Depth: 4 cm; Focal Depth: 2.4 cm; Mode: B

Table of Contents