Clarius Scanner HD3
revision 9 128
Clarius Scanner PAL HD3: Ocular (Ophthalmic) B-Mode, Linear
Array
Index Label MI TIS TIB TIC
At
Surface
Below
Surface
At
Surface
Below
Surface
Maximum Index Value 0.066 0.002 0.002 (a)
Index Component Value
0.002 0.002 0.002 0.002
Acoustic
Parameters
p
r,a
at Z
MI
(MPa) 0.211
P(mW) 0.043 0.043 #
P
1x1
(mW) 0.043 0.043
z
s
(cm) 2.00
z
b
(cm) 2.00
Z
MI
(cm) 2.00
Z
pii,a
(cm) 2.00
f
awf
(MHz) 10.265 10.265 10.265 #
Other Information prr (Hz) 4256
srr (Hz) 19.0
n
pps
2
I
pa,a
at Z
pii,a
(W/cm
2
)
1.788
I
spta,a
at Z
pii,a
or Z
sii,a
(mW/cm
2
)
0.030
I
spta,a
at Z
pii
or Z
sii
(mW/cm
2
)
0.124
p
r
at Z
pii
(MPa) 0.429
Operating Control
Conditions
Control 1
Control 2
Control 3
Control 4
Control 5
Control 6
Control 7
(a) This scanner is not intended for transcranial or neonatal cephalic use.
# Not reported for this operating condition; the maximum index value is not reported for the reasons shown in the Maximum
Index Value fields.
Control 1:Exam Type: Ocular; Depth: 4 cm; Focal Depth: 2 cm; Mode: B