Clarius Scanner HD3 Disinfecting
revision 9 38
For a list of compatible cleaners, see Cleaners & Disinfectants on page 74.
4. Repeat with new cleaning material if necessary.
5. Air-dry the Clarius Charger HD3. Alternatively, use a non-linting cloth.
Disinfecting
Disinfecting the Clarius Scanner HD3
Before you begin disinfecting, make sure you have cleaned the scanner (see Cleaning the
Clarius Scanner HD3 on page 36).
Disinfecting requires that you choose the proper disinfecting level. Determine the necessary
disinfection level by referring to Spaulding Classification on page 40. Once you have
determined the required disinfecting level, have the disinfectant ready and follow one of the
appropriate procedures below. Note that different levels of disinfection require different steps,
not just different solutions.
Intermediate Disinfection
Refer to Cleaners & Disinfectants on page 74 for a list of disinfectants recommended for
intermediate disinfection of the scanner.
If the scanner has come into contact with broken skin, mucosal membranes, or blood, it is
classified as semi-critical, and you must perform a high-level disinfection. See High-Level
Disinfection on page 38 for steps.
To disinfect your scanner (intermediate):
1. Make sure the fan is still detached from the scanner.
2. Disinfect the scanner by wiping with a cloth moistened with a compatible disinfectant.
Alternatively, use a premoistened disinfectant wipe. Use a swab for unreachable
areas.
3. Air-dry. Alternatively, use a non-linting cloth.
4. Examine the scanner for damage, such as cracks or splitting where fluid can enter. If
damage is evident, do not use the scanner and contact Clarius Mobile Health.
High-Level Disinfection
Refer to Cleaners & Disinfectants on page 74 for a list of disinfectants recommended for high-
level disinfection of the scanner.