Clarius Scanner HD3
revision 9 127
Clarius Scanner PAL HD3: Needle Enhance B-Mode, Linear
Array
Index Label MI TIS TIB TIC
At
Surface
Below
Surface
At
Surface
Below
Surface
Maximum Index Value 0.317 0.056 0.056 (a)
Index Component Value
0.056 0.056 0.056 0.056
Acoustic
Parameters
p
r,a
at Z
MI
(MPa) 0.901
P(mW) 1.464 1.464 #
P
1x1
(mW) 1.464 1.464
z
s
(cm) 2.456
z
b
(cm) 2.456
Z
MI
(cm) 2.456
Z
pii,a
(cm) 2.456
f
awf
(MHz) 8.103 8.103 8.103 #
Other Information prr (Hz) 5376
srr (Hz) 24.0
n
pps
2
I
pa,a
at Z
pii,a
(W/cm
2
)
40.987
I
spta,a
at Z
pii,a
or Z
sii,a
(mW/cm
2
)
1.053
I
spta,a
at Z
pii
or Z
sii
(mW/cm
2
)
4.079
p
r
at Z
pii
(MPa) 1.789
Operating Control
Conditions
Control 1 
Control 2
Control 3
Control 4
Control 5
Control 6
Control 7
(a) This scanner is not intended for transcranial or neonatal cephalic use.
# Not reported for this operating condition; the maximum index value is not reported for the reasons shown in the Maximum
Index Value fields.
Control 1: Exam Type: Vascular; Depth: 5.7 cm; Focal Depth: 2.5 cm; Mode: B

Table of Contents