Clarius Scanner HD3
revision 9 129
Clarius Scanner PAL HD3: PW Doppler Mode, Linear Array
Index Label MI TIS TIB TIC
At
Surface
Below
Surface
At
Surface
Below
Surface
Maximum Index Value 0.414 0.292 0.616 (a)
Index Component Value
0.292 0.135 0.616 0.616
Acoustic
Parameters
p
r,a
at Z
MI
(MPa) 1.067
P(mW) 9.236 9.236 #
P
1x1
(mW) 9.236 9.236
z
s
(cm) 1.681
z
b
(cm) 1.681
Z
MI
(cm) 1.681
Z
pii,a
(cm) 1.681
f
awf
(MHz) 6.630 6.630 6.630 #
Other Information prr (Hz) 8000
srr (Hz) —
n
pps
1
I
pa,a
at Z
pii,a
(W/cm
2
)
53.488
I
spta,a
at Z
pii,a
or Z
sii,a
(mW/cm
2
)
222.443
I
spta,a
at Z
pii
or Z
sii
(mW/cm
2
)
479.649
p
r
at Z
pii
(MPa) 1.567
Operating Control
Conditions
Control 1
Control 2
Control 3
Control 4
Control 5
Control 6
Control 7
(a) This scanner is not intended for transcranial or neonatal cephalic use.
# Not reported for this operating condition; the maximum index value is not reported for the reasons shown in the Maximum
Index Value fields.
— Not applicable for this scanner or mode.
Control 1: Exam Type: Vascular; Gate Depth: 4 cm; Focal Depth: 1.7 cm; Mode: PWD