Clarius Scanner HD3
revision 9 83
Clarius Scanner PA HD3
CLARIUS SCANNER PA HD3: B-MODE
Measurement Tolerance Range
Axial Distance < ± 2% 0 – 30.5 cm
Lateral Distance < ± 5% 0 – 30.5 cm
CLARIUS SCANNER PA HD3: M-MODE
Measurement Tolerance Range
Distance < ± 2% 0 – 30.5 cm
Time < ± 2% minimum = 0 ms
maximum = variable
1
1. Range depends on the viewing device used and the amount
of spectrum that can fit on the device's screen.
Heart Rate < ± 2% minimum = 1 beat
maximum = variable
a
CLARIUS SCANNER PA HD3: PW-MODE
Measurement Tolerance Range
Velocity < ± 2% 0 – 718.7 cm/s
Time < ± 2% minimum = 0 ms
maximum = variable
1
1. Range depends on the viewing device used and the amount
of spectrum that can fit on the device's screen.
Heart Rate < ± 2% minimum = 1 beat
maximum = variable
a
CLARIUS SCANNER PA HD3: DOPPLER SENSITIVITY
Sensitivity Depth (cm) Flow (ml/sec)
PW Doppler flow sensitivity at depth 5.0 0.1
CFI Doppler flow sensitivity at depth 5.0 0.1
PDI Doppler flow sensitivity at depth 5.0 0.1
PW Doppler depth sensitivity 9.4
CFI Doppler depth sensitivity 9.7
PDI Doppler depth sensitivity 10.2