Clarius Scanner HD3
revision 9 133
Clarius Scanner PAL HD3: PW Doppler Mode, Phased Array
Index Label MI TIS TIB TIC
At
Surface
Below
Surface
At
Surface
Below
Surface
Maximum Index Value 0.679 0.779 3.223 (a)
Index Component Value
0.779 0.497 2.570 3.223
Acoustic
Parameters
p
r,a
at Z
MI
(MPa) 0.893
P(mW) 94.638 94.638 #
P
1x1
(mW) 94.638 94.638
z
s
(cm) 3.785
z
b
(cm) 3.785
Z
MI
(cm) 3.785
Z
pii,a
(cm) 3.785
f
awf
(MHz) 1.728 1.728 1.728 #
Other Information prr (Hz) 7000
srr (Hz)
n
pps
1
I
pa,a
at Z
pii,a
(W/cm
2
)
25.554
I
spta,a
at Z
pii,a
or Z
sii,a
(mW/cm
2
)
429.727
I
spta,a
at Z
pii
or Z
sii
(mW/cm
2
)
673.272
p
r
at Z
pii
(MPa) 1.118
Operating Control
Conditions
Control 1 
Control 2
Control 3
Control 4
Control 5
Control 6
Control 7
(a) This scanner is not intended for transcranial or neonatal cephalic use.
# Not reported for this operating condition; the maximum index value is not reported for the reasons shown in the Maximum
Index Value fields.
— Not applicable for this scanner or mode.
Control 1: Exam Type: Cardiac; Gate Depth: 8 cm; Focal Depth: 3.8 cm; Mode: PWD