Clarius Scanner HD3
revision 9 85
Clarius Scanner PAL HD3 (Phased Array)
CLARIUS SCANNER PAL HD3: B-MODE, PHASED ARRAY
Measurement Tolerance Range
Axial Distance < ± 2% 0 – 30.5 cm
Lateral Distance < ± 5% 0 – 30.5 cm
CLARIUS SCANNER PAL HD3: M-MODE, PHASED ARRAY
Measurement Tolerance Range
Distance < ± 2% 0 – 30.5 cm
Time < ± 2% minimum = 0 ms
maximum = variable
1
1. Range depends on the viewing device used and the amount
of spectrum that can fit on the device's screen.
Heart Rate < ± 2% minimum = 1 beat
maximum = variable
a
CLARIUS SCANNER PAL HD3: PW-MODE, PHASED ARRAY
Measurement Tolerance Range
Velocity < ± 2% 0 – 718.7 cm/s
Time < ± 2% minimum = 0 ms
maximum = variable
1
1. Range depends on the viewing device used and the amount
of spectrum that can fit on the device's screen.
Heart Rate < ± 2% minimum = 1 beat
maximum = variable
a
CLARIUS SCANNER PAL HD3: DOPPLER SENSITIVITY, PHASED ARRAY
Sensitivity Depth (cm) Flow (ml/sec)
PW Doppler flow sensitivity at depth 5.0 0.1
CFI Doppler flow sensitivity at depth 5.0 0.1
PDI Doppler flow sensitivity at depth 5.0 0.1
PW Doppler depth sensitivity 9.4
CFI Doppler depth sensitivity 9.7
PDI Doppler depth sensitivity 10.2